A MOF-on-MOF heterostructure ratiometric/colorimetric dual-mode fluorescence sensor based on support vector machine for detecting tetracyclines in animal-derived foods

Zhehang Tian,Xieyang Zhang,Yuting Zhang,Zimeng Wu,Guanqun Luan,Luqian Bao,Yixin Ji,Mengyao Cui,Chunhua Li
DOI: https://doi.org/10.1016/j.foodchem.2024.140690
2024-12-01
Abstract:The misuse of tetracyclines in livestock production poses significant health risks. Thus, establishing convenient detection methods to replace complex laboratory tests for food safety is crucial. In this study, a heterostructure Zn-BTC/IRMOF-3 (denoted as ZBI) asynchronous response fluorescence sensor was developed for the qualitative and quantitative detection of tetracyclines in foods. The ZBI solution exhibited blue fluorescence under UV excitation; upon the introduction of tetracyclines, ZBI selectively recognized the tetracycline molecules through electron transfer, π-π stacking, and chelation, resulting in blue fluorescence quenching and green fluorescence enhancement. The ZBI sensor for tetracycline detection achieved recovery rates ranging from 93.91 to 111.91% in food samples, with a detection limit of as low as 0.086 μmol/L. Lastly, a portable sensing device using support vector classifier was constructed for detecting tetracyclines in real-life scenarios. Our findings introduce a new approach for fabricating fluorescence sensors and offer a novel method for detecting tetracyclines.
What problem does this paper attempt to address?