Slice-specific B1 + shimming improves the repeatability of multishot DWI at 7 T

Belinda Ding,Sydney Nicole Williams,Iulius Dragonu,Patrick Liebig,Sarah Allwood-Spiers,Paul McElhinney,Shajan Gunamony,Natasha Fullerton,David Andrew Porter
DOI: https://doi.org/10.1002/mrm.30208
2024-08-01
Abstract:Purpose: Compared with lower field strengths, DWI at 7 T faces the combined challenges of increased distortion and blurring due to B0 inhomogeneity, and increased signal dropouts due to B1 + inhomogeneity. This study addresses the B1 + limitations using slice-specific static parallel transmission (pTx) in a multi-shot, readout-segmented EPI diffusion imaging sequence. Methods: DWI was performed in 7 healthy subjects using MRI at 7 T and readout-segmented EPI. Data were acquired with non-pTx circular-polarized (CP) pulses (CP-DWI) and static pTx pulses (pTx-DWI) using slice-specific B1 + shim coefficients. Each volunteer underwent two scan sessions on the same day, with two runs of each sequence in the first session and one run in the second. The sequences were evaluated by assessing image quality, flip-angle homogeneity, and intrasession and intersession repeatability in ADC estimates. Results: pTx-DWI significantly reduced signal voids compared with CP-DWI, particularly in inferior brain regions. The use of pTx also improved RF uniformity and symmetry across the brain. These effects translated into improved intrasession and intersession repeatability for pTx-DWI. Additionally, re-optimizing the pTx pulse between repeat scans did not have a negative effect on ADC repeatability. Conclusion: The study demonstrates that pTx provides a reproducible image-quality increase in multishot DWI at 7 T. The benefits of pTx also extend to quantitative ADC estimation with regard to the improvement in intrasession and intersession repeatability. Overall, the combination of multishot imaging and pTx can support the development of reliable, high-resolution DWI for clinical studies at 7 T.
What problem does this paper attempt to address?