Application of Analytical Quality by Design to the development and validation of reduced and non-reduced capillary electrophoresis analytical procedures for mAb purity determination

Terezie Cernosek,Matthew Dalphin,Joseph Sakowski,Sue Behrens,Peter Wunderli
DOI: https://doi.org/10.1016/j.jpba.2024.116386
2024-10-15
Abstract:Capillary electrophoresis-sodium dodecyl sulfate (CE-SDS) is a common analytical procedure used to quantitate critical quality attributes relating to the purity and glycosylation of monoclonal antibodies (mAbs). In this study, the application of an Analytical Quality by Design framework incorporating Design of Experiments was used to develop and validate both non-reduced (CE-NR) and reduced (CE-R) versions of this analytical procedure. Formal risk assessments were used to identify critical method attributes for optimization based on their potential impacts to performance criteria outlined in an analytical target profile. The resulting response surfaces connecting these critical factors to method performance were then utilized to generate a harmonized procedure to reduce execution risk across CE-R and CE-NR applications. Validation of these procedures according to regulatory guidelines support that they meet their required performance criteria, and a multivariate assessment of procedure robustness indicates that method parameters are in a sufficient state of control to ensure appropriate quantitation of mAb quality. Overall, this study demonstrates the utility of adopting an Analytical Quality by Design framework to leverage multidimensional knowledge from multiple critical method parameters to ensure an analytical procedure is fit-for-purpose.
What problem does this paper attempt to address?