Multiple site readings from a transcutaneous bilirubinometer

D. Tudehope,Alan Chang
DOI: https://doi.org/10.1111/j.1440-1754.1982.tb01999.x
1982-06-01
Abstract:ABSTRACT. The Transcutaneous Bilirubinometer was evaluated on 95 occasions on 66 infants who were clinically jaundiced but not receiving phototherapy. Meter readings were obtained from seven different body sites within 30 minutes of sampling blood for serum bilirubin estimation. A cephalocaudal progression of dermal icterus was demonstrated. Meter readings from forehead, sternum, thigh, upper and lower back correlated closely with serum bilirubin (r ≧ .93). The 95% confidence limits for forehead and sternum were ± 52.8 μmol/l. Multiple regression analysis using all 7 body sites increased the correlation coefficient to .97 and decreased the confidence limits to 44.4 μmol/l. However, we would recommend the forehead and sternum as the most reliable sites for meter readings with acceptable accuracy and precision. Meter readings from preterm and low birth weight babies tend to over‐estimate serum bilirubin. Further Australian studies are necessary to establish regression lines for infants of different racial backgrounds.