The design of wedge filters for X-ray therapy.

A. Groom
DOI: https://doi.org/10.1259/0007-1285-24-288-676
1951-12-01
British Journal of Radiology
Abstract:An analysis of the dosage distributions produced in a wax phantom by copper wedge filters of the plateau type has been made for a wide range of wedge patterns and for different field sizes at 226 kV. General formulae are given for the construction of wedge filters for use between 180—250 kV, giving, for various isodose slopes, the maximum output consistent with satisfactory isodose configuration. The advantages of the plateau type of wedge filter over filters of a pure wedge section are discussed.
What problem does this paper attempt to address?