Visual and quantitative lateral flow immunoassay based on Au@PS SERS tags for multiplex cardiac biomarkers

Li Wang,Jianli Sun,Xiaoxuan Wang,Milan Lei,Zengliang Shi,Lei Liu,Chunxiang Xu
DOI: https://doi.org/10.1016/j.talanta.2024.126040
IF: 6.1
2024-07-01
Talanta
Abstract:Rapid and sensitive detection of multiple biomarkers by lateral flow immunoassay (LFIA) remains challenging for signal amplification for commonly used nanotags. Herein, we report a novel LFIA strip for visual and highly sensitive analysis of two cardiac biomarkers based on functionalized gold nanoparticles @ polystyrene microsphere (Au@PS)microcavity as surface-enhanced Raman scattering (SERS) tags. Antibody-modified Au@PS was designed as a SERS label. The evanescent waves propagating along the surface of the PS microcavity and the localized surface plasmons of the gold nanoparticles were coupled to enhance the light-matter interaction synergistically for Raman signal enhancement. In this strategy, the proposed Au@PS SERS tags-based LFIA was carried out to quantify the content of the heart failure and infarct biomarkers synchronously within 15 min and get the limits of detection of 1 pg/mL and 10 pg/mL for cardiac troponin I (cTnI) and N-terminal natriuretic peptide precursor (NT-proBNP), respectively. The results demonstrated 10-20 folds more sensitivity than that of the standard colloidal gold strip and fluorescent strip for the same biomarkers. This novel quantitative LFIA shows promise as a high-sensitive and visual sensing method for relevant clinical and forensic analysis.
What problem does this paper attempt to address?