Enhancing Guidewire Efficacy for Transradial Access: The EAGER Randomized Controlled Trial
Adam C Bland,William Meere,Philopatir Mikhail,Eunice Chuah,Eleanor Redwood,David Ferreira,Nicklas Howden,Adam Perkovic,Samantha L Saunders,Amy Kelty,Tony Kull,Andrew Hill,Roberto Spina,Kiran Sarathy,Austin May,Michael Parkinson,Mark Ishak,Nicholas Collins,Andrew Boyle,Maged William,Prajith Jeyaprakash,Tom J Ford
DOI: https://doi.org/10.1161/CIRCINTERVENTIONS.124.014529
Abstract:Background: The 1.5 mm Baby J hydrophilic narrow J-tipped wire is a development of the standard 0.035" 3-mm J-tipped peripheral guidewire, designed to improve efficiency of transradial coronary procedures by safely navigating small caliber radial arteries to the aorta. There is currently a lack of evidence comparing the procedural success and safety of different peripheral guidewires used in transradial cardiac procedures. We compared the efficacy and safety of a narrow J-tipped hydrophilic 0.035" wire (intervention, Radifocus Baby J guidewire; TERUMO Co, Tokyo, Japan) versus a standard fixed-core 0.035" J wire (control). Methods: Investigator-initiated, blinded, Australian, multicenter randomized trial in patients undergoing clinically indicated coronary angiography or percutaneous coronary intervention. Patients were randomized 1:1 to use either the control guidewire or the intervention guidewire. The primary end point (technical success) was defined as gaining aortic root access with the randomized guidewire. Results: In all, 330 patients were randomized between October 2022 and June 2023 (median age was 69 years, 36% were female, and body mass index was 29 kg/m2). The primary end point was achieved more frequently in the intervention group (96% versus 84%; absolute risk reduction 12% [95% CI, 5.7-18.3]; P<0.001). Women assigned to the control wire experienced a higher failure rate compared with men (31% versus 8% in men; P<0.001). Fluoroscopy time was significantly lower in the Baby J group (median, 344 versus 491 seconds; P=0.024). The main mechanisms of failure using the control wire were radial artery spasm (15/26; 57%) and subclavian tortuosity (5/26; 19.2%). There were no differences in overall procedure times, major adverse cardiovascular events, or vascular complications between guidewires. Conclusions: A narrow 1.5 mm J-tipped hydrophilic guidewire resulted in greater technical success and reduced fluoroscopy time compared with the standard 3-mm J-tipped nonhydrophilic guidewire. The guidewire is safe and demonstrated key incremental benefits for the transradial approach, particularly in women. Registration: URL: https://www.anzctr.org.au/; Unique identifier: ACTRN12622001557729.