Tumpy, the Camel, and the Electrical Dose Concept for Ventricular Defibrillation

L. Geddes
DOI: https://doi.org/10.1046/j.1460-9592.2003.t01-1-00223.x
2003-07-01
Abstract:In 1954, the Baylor University College of Medicine physiology department, under the chairmanship of the late Hebbel Hoff, started teaching direct heart defibrillation to all first-year medical students using the dog. Each student was required to induce fibrillation, learn the art of effective cardiac compression, and then defibrillate successfully with 60-Hz current applied to transventricular electrodes (salad spoons). In 1947 Beck et al.1 first successfully defibrillated the human heart with a 0.1-second train of 60-Hz power line current applied to transventricular electrodes. Prior to 1947, clinicians achieved defibrillation by first perfusing the coronary arteries with KCl (to achieve depolarization), and then perfusing them with CaCl2 (to repolarize the myocardial fibers). The simpler, easy-to-use electrical method was soon adopted by clinicians. But how much current was needed? How long should it flow, and what is the effect of heart size? Regarding the effect of current pulse duration,2 showed that the Lapicque strength-duration curve applied to defibrillation with a single capacitor discharge current pulse. Koning et al.3 showed the same using rectangular wave pulses. The results of our study4 showed that the law applies for damped sinusoidal pulses. Thus, it was shown that a train of 60 Hz current pulses was not needed; a single pulse is adequate.