Abstract:Blood pressure (BP) is one of the four main vital signs in medicine and may be a useful signal for wellness tracking and for user-aware interfaces in human-computer interaction. The current standard for BP measurement uses cuff-based devices that block an artery temporarily to get a single, discrete measurement of BP. Recently, there have been significant efforts to measure correlates of BP continuously and non-invasively from relevant signals like photoplethysmography (PPG), which responds to volumetric changes in arteries due to blood pulsations. In this paper, we explore a novel setup with two points of instrumentation, one on the head and a second on the wrist, for recording PPG and estimating the pulse wave velocity, which is a major correlate of BP, along with other waveform-related features. We prospectively tested the device on 10 subjects who followed a protocol for the deliberate variation of BP while ground truth measurements were taken using a reference cuff-device. Generic absolute BP models, which use the collected data for leave-one-subject-out cross-validation, yielded an error of -0.14 ± 7.3 mmHg for systolic BP (SBP) and -0.21±6.7 mmHg for diastolic BP (DBP), which are within the regulatory limits of 5 ± 8 mmHg. Notably, the correlation between the predicted BPs and the ground truth BPs was higher for SBP (r = 0.74, p < 0.001) than for DBP (r = 0.34, p < 0.001). The results show that the proposed form factor can extract BP-related features that could be used for continuous, cuff-less BP monitoring.