Measurement of the neutron response of a graphite-CO2 ionization chamber.

J. Sayeg,P. Harris,T. A. Allison
DOI: https://doi.org/10.1148/70.1.95
IF: 19.7
Radiology
Abstract:In the determination of neutron and gamma tissue doses by the tissue-equivalent ionization chamber method (1), the neutron response of the gamma-measuring graphite-CO2 ionization chamber has been calculated from first collision theory with the assumption that the response is mainly due to the CO2 gas (2). Experiments have been conducted with monoenergetic neutrons from the Los Alamos Van de Graaff generator to determine this response as a function of neutron energy. The detectors employed were a Hurst proportional counter, a beryllium-shelled condenser-type tissue-equivalent ionization chamber, and a beryllium-shelled condenser-type graphite-CO2 chamber (3). The neutron response, K, was calculated from the relation: where a is the efficiency of our tissue-equivalent chamber in measuring a neutron dose when the chamber is calibrated with hard x-rays or gamma rays, T is the dose in r/n/cm2 (in terms of an x-ray calibration) measured by the tissue-equivalent ionization chamber, G is the dose in r/n/cm2 (in t...