Validity of the Entralpi force plate in the assessment of finger flexor performance metrics in rock climbers

Keenan B MacDougall,Zachary J McClean,Brian R MacIntosh,Jared R Fletcher,Saied J Aboodarda
DOI: https://doi.org/10.1080/14763141.2023.2259356
2023-09-18
Abstract:This study assessed the validity of the Entralpi force plate in the assessment of finger flexor performance in rock climbers. In addition to a static force evaluation, peak force, peak impulse, and total impulse were measured during 30 all-out performance trials by 15 participants, in which force during the trials was recorded simultaneously by the Entralpi and a Pasco force plate. Agreement between devices was assessed by a variety of statistical analyses, including intraclass correlation coefficient (ICC), coefficient of variation (CV), and Bland-Altman analyses. The static force evaluation showed a mean relative error of 0.41% and excellent day-to-day reliability (ICC = 1; CV = 0.03%). Peak force, peak impulse, and total impulse from the performance trials demonstrated strong agreement (ICC ≥ 0.991, CV ≤ 1.9%, Bland-Altman mean bias ≤ 0.5%). These results illustrate that the Entralpi force plate provides accurate and reliable data for rock climbing related tasks at an affordable cost.
What problem does this paper attempt to address?