Comparison of a new implantable collamer lens formula to standards formulas using spectral domain optical coherence tomography

Alaa Eldanasoury,Harkaran Bains,Stefan Pieger
DOI: https://doi.org/10.1007/s10792-023-02861-5
Abstract:Purpose: To use spectral-domain optical coherence tomography (SD-OCT) data to develop a new implantable collamer lens (ICL) sizing formula and compare vault outcomes with the Online Calculation and Ordering System™ (OCOS) and the NK2 formula. Methods: Consecutive eyes (n = 237) were evaluated that had undergone ICL/toric ICL implantation. Actual ICL vaults were measured, and a what-if analysis was performed to predict vault values with the NK2 formula using SD-OCT data. To develop a new formula (EPB), multiple regression analysis was performed with different parameters than the NK2 formula. Predicted vaults with NK2 and EPB formulas were compared to the actual vaults. Results: Parameters that were correlated with optimal ICL size were white-to-white, anterior chamber width, lens rise and desired refractive correction. The mean postoperative vault was 489 ± 258 μm. At last visit, 94.5% of eyes were within the manufacturer's acceptable vault range. Predicted vaults in the acceptable range were 74 and 87% with the NK2 and EPB formulas, respectively. Six percent had a predicted vault less than 100 μm with the EPB formula compared to 1% for actual outcomes. The NK2 formula resulted in a shift toward higher predicted vaults while the EPB formula was similar to the actual postoperative vaults but with slightly more cases with extremely low and high vaults. Conclusion: SD-OCT data with OCOS result in good postoperative vaults. Further refinement is required to the NK2 for use with SD-OCT data. Although the EPB formula provides acceptable predicted vaults, further refinement with a larger sample size is needed.
What problem does this paper attempt to address?