Impact on collateral flow of devices used for endovascular treatment of stroke: an in-vitro flow model

Manuel Requena,Jiahui Li,Riccardo Tiberi,Pere Canals,Marta Olive Gadea,Marta de Dios Lascuevas,Magda Jabłońska,Judith Cendrero,Alvaro Garcia-Tornel,Alejandro Tomasello,Marc Ribo
DOI: https://doi.org/10.1136/jnis-2023-020602
2024-09-17
Abstract:Background: Collateral blood supply of distal vessels has been linked to clinical outcome, infarct volume and recanalization rates in patients with large vessel occlusion. Our study aimed to explore the effects of catheterization during mechanical thrombectomy in collaterals. Methods: We quantified the flow diversion effect secondary to arterial occlusions in an in vitro model which was connected in a flow-loop setup with a saline reservoir and a pump supplying pulsatile flow. Clot analogs were embolized to the middle cerebral artery (MCA) M1 or M2 segments. We used the same model with a clamped anterior communicating artery (AComA) to simulate its absence. An ultrasound flow sensor was placed at the vessel of interest. Flow rates and pressures were evaluated according to the following catheter locations: baseline (1) before and (2) after the occlusion; (3) 8F guiding catheter at the internal carotid artery (ICA) bulb; (4) at the cavernous segment; (5) at the cavernous segment a 0.071" distal access catheter at proximal M1; (6) 8F balloon guide catheter inflated. Results: Collateral blood flow measured at distal anterior cerebral artery (ACA) (M1-MCA occlusion) and M2-MCA (M2-MCA occlusion) was progressively reduced as catheters were advanced through the ICA and MCA. In the lacking AComA model, the flow was further diminished as compared with the model with a patent AComA. Conclusion: Our in vitro study showed a progressive reduction of collateral blood flow due to the advance of catheters during mechanical thrombectomy.
What problem does this paper attempt to address?