A Novel ELISA for the Quantification of Serum Levels of 2 Glycosylated Variants of Apolipoprotein J: Biomarkers for Myocardial Ischemia

Lola Fernández Encinas,Nuria Lluch,Alan H B Wu,Juan Carlos Kaski,Lina Badimon,Judit Cubedo
DOI: https://doi.org/10.1093/jalm/jfad034
2023-09-07
Abstract:Background: Previous studies have pointed out a potential role of ApoJ-Glyc as a biomarker of cardiac ischemia. The aim of this study was to validate the analytical performance of 2 novel ELISAs against 2 different glycosylated ApoJ variants, ApoJ-GlycA2 and ApoJ-GlycA6. Methods: The analytical measuring range, limit of blank (LoB), lower limit of quantification (LoQ), precision, accuracy, recovery, cross-reactivity, and stability were evaluated in serum samples. Results: The analytical measuring range was 500-16 000 ng/mL for ApoJ-GlycA2 and 125-4000 ng/mL for ApoJ-GlycA6, with a LoB of 455 ng/mL and 121 ng/mL for ApoJ-GlycA2 and ApoJ-GlycA6, respectively. The LoQ was 500 ng/mL for ApoJ-GlycA2 and 125 ng/mL for ApoJ-GlycA6. The assay performance fulfills the acceptance criteria established in the European Medicines Agency Guideline on bioanalytical method validation. Specifically, the calibration range variability is <15% for ApoJ-GlycA2 and ApoJ-GlycA6; the accuracy is <15% for ApoJ-GlycA2 and ApoJ-GlycA6; the between- and within-run precision is <15% for ApoJ-GlycA6 and ≤20% for ApoJ-GlycA2; and the total allowable error is <30% for ApoJ-GlycA2 and ApoJ-GlycA6. Cross-reactivity studies revealed the absence of cross-reactivity with endogenous components of the matrix (using ApoJ-depleted serum), with nonglycosylated ApoJ and with transferrin (as a high abundant N-glycosylated serum protein). Both ApoJ-GlycA2 and ApoJ-GlycA6 measurements were stable after storage of serum samples at -80°C for 24 months. Conclusions: The newly developed ELISAs to quantify ApoJ-GlycA2 and ApoJ-GlycA6 serum levels showed an acceptable analytical performance according to European Medicines Agency guidelines on bioanalytical method validation in terms of precision, accuracy, recovery, cross-reactivity, and stability.
What problem does this paper attempt to address?