The value of urine sediment analyzer in screening white blood cells and bacteria for urinary tractinfection
Li-li GUO,Kui ZHANG
DOI: https://doi.org/10.3969/j.issn.1672-271X.2017.03.003
2017-01-01
Abstract:Objective To evaluate the performance of white blood cells (WBC) and bacteria (BACT) analyses in urinalysis analyzer Uf-1000i and determine the potential application for urinary tract infection screening.Methods A total of 304 cases with suspected urine infection were collected.The performance including precision, linearity and carryover contamination rates in Uf-1000i were determined.ROC curve was determined, and the sensitivity, specificity, false negative rates, expected positive rate and expected negative rate were evaluated compared with gold standard method-urine culture.Results Uf-1000i showed good performance of within-run precision, intra-day precision, carryover, and linearity range for both WBC and BACT.The cutoff value based on ROC analysis for urinary infection using Uf-1000i was further determined: 173.5 WBC/μL and 513.5 bacteria/μL.The sensitivity, specificity, false positive rate, false negative rate, positive predictive value, negative predictive for either WBC or BACT in Uf-1000i were 83.3%, 73.0%, 27.0%, 16.7%, 62.9%, 88.8%, respectively.The sensitivity, specificity, false positive rate, false negative rate, positive predictive value, negative predictive for both WBC and BACT in Uf-1000i were 54.6%,93.9%, 6.1%, 45.4%, 83.1%, 79.0%, respectively.Conclusion The performance of WBC and BACT analysis in Uf-1000i fulfilled the requirement.Therefore, it was able to be applied as rapid screening equipment for clinical urinary infection, which facilitates identification of bacteria species and offers guidance of early drug administration during clinical practice.