Commutability evaluation of candidate reference materials and ERM‐DA470k/IFCC for immunoglobulin M using two international approaches
Lu Sang,Chunying Wu,Huijuan Chen,Wei Liu,Dawei Huang,Xi Yang,Xinrui Guo,Ruifang Cui,Ning Wang,Rui Zhang,Yuhong Yue,Hong Guo,Minghao Wang,Yutong Miao,Qingtao Wang,Shunli Zhang
DOI: https://doi.org/10.1002/jcla.24955
IF: 3.124
2023-08-13
Journal of Clinical Laboratory Analysis
Abstract:Our study showed that commutability of immunoglobulin M (IgM) frozen pooled human serum (PHS) was not so good. High concentration of PHS were commutable on all six detection systems using the CLSI approach. Low and medium concentration of PHS showed unsatisfied commutability. High concentration of IgM pure diluted materials (HPDM), which was diluted in healthy human serum have good commutability, has the potential to become reference materials. ERM‐DA470k/IFCC showed different commutability in different matrix. Background This study aimed to assess the commutability of frozen pooled human serum (PHS), high concentration of Immunoglobulin M (IgM) pure diluted materials (HPDM), commercialized pure materials (CPM), and dilutions of ERM‐DA470k/IFCC in IgM detection using the CLSI and IFCC approaches, to support standardization or harmonization of IgM measurement. Methods Twenty‐four serum samples, relevant reference materials (PHS, HPDM, CPM), and different ERM‐DA470k/IFCC dilutions were analyzed in triplicate using six routine methods. The commutability of the relevant reference materials was carried out following CLSI EP30‐A and IFCC bias analysis. Results According to the CLSI approach, low, medium, and high concentrations of PHS, HPDM, and CPM were commutable on 10, 13, 15, 13, and 8 of 15 assay combinations, respectively. Using the IFCC approach, low, medium, and high concentrations of PHS, HPDM, and CPM were commutable on 10, 11, 9, 15, and 10 of 15 assay combinations, respectively. The ERM‐DA470k/IFCC dilutions with D‐PBS and RPMI‐1640 Medium were commutable on 13 of 15 assay combinations according to CLSI and were commutable on all 15 assay combinations using IFCC approach. Conclusions High concentration of PHS were commutable on all six detection systems using the CLSI approach. Low and medium concentration of PHS showed unsatisfied commutability. HPDM, not CPM have good commutability, has the potential to become reference materials. ERM‐DA470k/IFCC diluted with different medium showed different commutability.
medical laboratory technology