Photographic recorder for erythrocyte sedimentation rates.

O. Gyde
DOI: https://doi.org/10.1136/jcp.25.1.96
1972-01-01
Journal of Clinical Pathology
Abstract:A photographic recorder for erythrocyte sedimentation rates (ESR) was developed and has now been in routine use in this laboratory for over a year. The need for such a device arose from the difficulty in dealing with specimens arriving too late in the day for the technical staff to read the results. As many of the requests involved outpatients attending clinics it was thought uneconomic in terms of both hospital and patient to bring them back for this single test. Attempts to read ESRs on an overtime or emergency basis proved to be either expensive or unsuccessful and keeping the specimens refrigerated overnight produced unreliable results. The prototype apparatus was constructed from various spare parts and iss shown in the photograph and drawing (Figs. 1 and 2). It consists essentially ofa sturdy box with a removable rack for Westergren tubes at one end and a Polaroid camera at the other. The camera shutter is operated by a solenoid switch, Fig. I. which is itself set off by a delay timer after an interval of 60 minutes. The latter switch also controls the light source which is turned on just before the solenoid is activated and the shutter opened. The exposed film can be removed from the camera / IO\ at any convenient time, which is usually after 18 to 48 hours, and is simply developed on the spot / LAYS