Real-world clinical outcomes with a next-generation left atrial appendage closure device: the FLXibility Post-Approval Study

Tim R Betts,Marek Grygier,Jens Erik Nielsen Kudsk,Thomas Schmitz,Marcus Sandri,Gavino Casu,Martin Bergmann,David Hildick-Smith,Thomas Christen,Dominic J Allocco,FLXibility investigators
DOI: https://doi.org/10.1093/europace/euac270
2023-03-30
Abstract:Aims: The FLXibility Post-Approval Study collected data on unselected patients implanted with a WATCHMAN FLX in a commercial clinical setting. Methods and results: Patients were implanted with a WATCHMAN FLX per local standard of care, with a subsequent first follow-up visit from 45 to 120 days post-implant and a final follow-up at 1-year post-procedure. A Clinical Event Committee adjudicated all major adverse events and TEE/CT imaging results were adjudicated by a core laboratory. Among 300 patients enrolled at 17 centres in Europe, the mean age was 74.6 ± 8.0 years, mean CHA2DS2-VASc score was 4.3 ± 1.6, and 62.1% were male. The device was successfully implanted in 99.0% (297/300) of patients. The post-implant medication regimen was DAPT for 87.3% (262/300). At first follow-up, core-lab adjudicated complete seal was 88.2% (149/169), 9.5% (16/169) had leak <3 mm, 2.4 (4/169) had leak ≥3 mm to ≤5 mm, and 0% had >5 mm leak. At 1 year, 93.3% (280/300) had final follow-up; 60.5% of patients were on a single antiplatelet medication, 21.4% were on DAPT, 5.6% were on direct oral anticoagulation, and 12.1% were not taking any antiplatelet/anticoagulation medication. Adverse event rates through 1 year were: all-cause death 10.8% (32/295); CV/unexplained death 5.1% (15/295); disabling and non-disabling stroke each 1.0% (3/295, all non-fatal); pericardial effusion requiring surgery or pericardiocentesis 1.0% (3/295); and device-related thrombus 2.4% (7/295). Conclusion: The WATCHMAN FLX device had excellent procedural success rates, high LAA seal rates, and low rates of thromboembolic events in everyday clinical practice.
What problem does this paper attempt to address?