Evaluation of two methods for measuring delta-aminolaevulinic acid in urine. II. The influence of some Ehrlich-positive' substances on the measurements.

B. Haeger‐Aronsen,M. Abdulla
DOI: https://doi.org/10.3109/00365517209081077
1972-04-01
Scandinavian Journal of Clinical and Laboratory Investigation
Abstract:Published results and earlier personal findings made on comparison between the determinations of δ-aminolaevulinic acid (ALA) in urine with the original and specific but time-consuming American method by Mauzerall & Granick and its recent simplified Polish modification by Grabecki et al. were confirmed. Thus, the latter method proved reliable, quick and cheap. But since it gives higher values than the American method, evaluation of the results must be revised. New normal values and limits of different ranges in the control of lead workers are suggested. Since Grabecki's modification does not eliminate disturbing substances, porfobilinogen (PBG), urobilinogen (UBG), p-aminosalicylic acid (PAS), and sulfonamides, all Ehrlich-positive compounds, were studied for any influence on the results. The presence of PBG was found to result in false high values, while UBG, PAS, and sulfonamides had no effect on the analysis of ALA in the urine with the method of Grabecki et al. Mauzerall & Granick's method is therefor...