Life in an optical fiber: Monitoring of cell cultures with microcavity in-line Mach-Zehnder interferometer

Monika Janik,Malwina Sosnowska,Tomasz Gabler,Marcin Koba,Anna Myśliwiec,Marta Kutwin,Ewa Sawosz Chwalibóg,Mateusz Śmietana
DOI: https://doi.org/10.1016/j.bios.2022.114718
2022-12-01
Abstract:Monitoring cell adhesion and growth are crucial for various applications involving drug screening, cytotoxicity, and cytocompatibility studies. However, acquiring accurate information about the growing state and responsiveness to a treatment of a cell system in a real-time and label-free manner is still a challenge. This work presents the first research on direct, real-time, and label-free adherent cell culture monitoring using a microcavity in-line Mach-Zehnder interferometer (μIMZI) fabricated in an optical fiber. The sensing solution based on μIMZI offers a great advantage over many other monitoring concepts tracking the changes taking place on the microcavity's bottom surface and within its volume, thus offering a greater penetration depth. In this study, we verified performance of the approach using a non-cancer bone marrow stromal cell line HS-5. The results demonstrate that the changes of the acquired signal are closely related to the different states of cells' adhesion, proliferation, morphology, and variation of mass. Thus, this label-free, real-time μIMZI-based monitoring technique gives a great promise to the analysis or monitoring of relevant new treatments in future scientific, as well as clinical applications.
What problem does this paper attempt to address?