Proximity effect thermometer for local temperature measurements on mesoscopic samples

J. Aumentado,J. Eom,V. Chandrasekhar,P.M. Baldo,L.E. Rehn
DOI: https://doi.org/10.48550/arXiv.cond-mat/0105083
2001-05-04
Mesoscale and Nanoscale Physics
Abstract:Using the strong temperature dependent resistance of a normal metal wire in proximity to a superconductor, we have been able to measure the local temperature of electrons heated by flowing a dc current in a metallic wire to within a few tens of millikelvin at low temperatures. By placing two such thermometers at different parts of a sample, we have been able to measure the temperature difference induced by a dc current flowing in the sample. This technique may provide a flexible means of making quantitative thermal and thermoelectric measurements on mesoscopic metallic samples.