Comparison of Measured and Calculated Dose Rates Near Nuclear Medicine Patients
Y. Yi,M. G. Stabin,M. H. McKaskle,M. D. Shone,A. B. Johnson
DOI: https://doi.org/10.1097/hp.0b013e318290cc0e
2013-01-01
Health Physics
Abstract:Widely used release criteria for patients receiving radiopharmaceuticals (NUREG-1556, Vol. 9, Rev. 1, Appendix U) are known to be overly conservative. The authors measured external exposure rates near patients treated with I-131, Tc-99m, and F-18 and compared the measurements to calculated values using point and line source models. The external exposure dose rates for 231, 11, and 52 patients scanned or treated with 131I, Tc-99m, and F-18, respectively, were measured at 0.3 m and 1.0 m shortly after radiopharmaceutical administration. Calculated values were always higher than measured values and suggested the application of "self-shielding factors," as suggested by Siegel et al. in 2002. The self-shielding factors of point and line source models for I-131 at 1 m were 0.60 +/- 0.16 and 0.73 +/- 0.20, respectively. For Tc-99m patients, the self-shielding factors for point and line source models were 0.44 +/- 0.19 and 0.55 +/- 0.23, and the values were 0.50 +/- 0.09 and 0.60 +/- 0.12, respectively, for F-18 (all FDG) patients. Treating patients as unshielded point sources of radiation is clearly inappropriate. In reality, they are volume sources, but treatment of their exposures using a line source model with appropriate self-shielding factors produces a more realistic, but still conservative, approach for managing patient release.