Exploring the use of infrared absorption spectroscopy and two-trace two-dimensional correlation analysis for the resolution of multi-component drug mixtures

Lea Gozdzialski,Bruce Wallace,Isao Noda,Dennis Hore
DOI: https://doi.org/10.1016/j.saa.2022.121684
2022-12-05
Abstract:Community drug checking provides an essential service that responds to the unpredictable and variable supply of illicit drugs. Point of care detection of trace components using portable infrared spectrometers is a harm reduction measure to prevent overdose. This study investigates the ability of weighted subtraction and two-trace two-dimensional (2T2D) correlation analysis to reveal the presence of heroin in an opioid mixture that contains heroin and fentanyl mixed with caffeine as a cutting agent. In both methods, a spectral trace was identified that provided reasonably high correlation scores to heroin when compared to entries in drug libraries. The two-trace correlation analysis produced a higher match score, suggesting that future improvements in spectral unmixing methods may enhance the reliability of detecting trace components in drugs.
What problem does this paper attempt to address?