Validation and Verification of High-Fidelity Simulations of Thoracic Stent-Graft Implantation

Anna Ramella,Francesco Migliavacca,Jose Felix Rodriguez Matas,Frederic Heim,Francesca Dedola,Stefania Marconi,Michele Conti,Sara Allievi,Tim J Mandigers,Daniele Bissacco,Maurizio Domanin,Santi Trimarchi,Giulia Luraghi
DOI: https://doi.org/10.1007/s10439-022-03014-y
Abstract:Thoracic Endovascular Aortic Repair (TEVAR) is the preferred treatment option for thoracic aortic pathologies and consists of inserting a self-expandable stent-graft into the pathological region to restore the lumen. Computational models play a significant role in procedural planning and must be reliable. For this reason, in this work, high-fidelity Finite Element (FE) simulations are developed to model thoracic stent-grafts. Experimental crimp/release tests are performed to calibrate stent-grafts material parameters. Stent pre-stress is included in the stent-graft model. A new methodology for replicating device insertion and deployment with explicit FE simulations is proposed. To validate this simulation, the stent-graft is experimentally released into a 3D rigid aortic phantom with physiological anatomy and inspected in a computed tomography (CT) scan at different time points during deployment with an ad-hoc set-up. A verification analysis of the adopted modeling features compared to the literature is performed. With the proposed methodology the error with respect to the CT is on average 0.92 ± 0.64%, while it is higher when literature models are adopted (on average 4.77 ± 1.83%). The presented FE tool is versatile and customizable for different commercial devices and applicable to patient-specific analyses.
What problem does this paper attempt to address?