Quartz-enhanced photoacoustic NH3 sensor exploiting a large-prong-spacing quartz tuning fork and an optical fiber amplifier for biomedical applications

Zhijin Shang,Shangzhi Li,Biao Li,Hongpeng Wu,Angelo Sampaolo,Pietro Patimisco,Vincenzo Spagnolo,Lei Dong
DOI: https://doi.org/10.1016/j.pacs.2022.100363
IF: 9.656
2022-05-04
Photoacoustics
Abstract:A sensor system for exhaled ammonia (NH3) monitoring exploiting quartz-enhanced photoacoustic spectroscopy (QEPAS) was demonstrated. An erbium-doped fiber amplifier (EDFA) with an operating frequency band targeting an NH3 absorption line falling at 1531.68 nm and capable to emit up to 3 W of optical power was employed. A custom T-shaped grooved QTF with prong spacing of 1 mm was designed and realized to allow a proper focusing of the high-power optical beam exiting the EDFA between the prongs. The performance of the realized sensor system was optimized in terms of spectrophone parameters, laser power and modulation current, resulting in a NH3 minimum detectable concentration of 14 ppb at 1 s averaging time, corresponding to a normalized noise equivalent absorption coefficient (NNEA) of 8.15 × 10-9 cm-1 W/√Hz. Continuous measurements of the NH3 level exhaled by 3 healthy volunteers was carried out to demonstrate the potentiality of the developed sensor for breath analysis applications.
What problem does this paper attempt to address?