Multicenter Prospective Study to Validate a New Transient Elastography Device for Staging Liver Fibrosis in Patients with Chronic Hepatitis B
Wei Jia Duan,Xiao Zhong Wang,An Lin Ma,Jia Shang,Yue Min Nan,Zhi Liang Gao,Hong Tang,Qing Chun Fu,Qing Xie,Qing Mao,Jun Qi Niu,Tao Han,Jun Li,Ying Han,Jian Biao Cao,Yuan,Xiao Yan Shi,Fu Dong Lv,Tai Ling Wang,Hong Ma,Hong You,Xiao Juan Ou,Ji Dong Jia
DOI: https://doi.org/10.1111/1751-2980.12924
2020-01-01
Journal of Digestive Diseases
Abstract:ObjectivesTo validate the operational and diagnostic performances of a new device for transient elastography (TE), FibroTouch, for liver fibrosis in patients with chronic hepatitis B (CHB).MethodsIn this prospective multicenter study, adult patients with CHB and valid liver pathological results were recruited to validate the operational and diagnostic performance of a TE device by FibroTouch for staging liver fibrosis.ResultsIn total, 517 patients with histologically proven CHB were enrolled. All had achieved at least 10 successful liver stiffness measurements (LSM), resulting in a success rate of 99.1% and reliable evaluations of 95.2%. Altogether 412 patients were included to analyze the diagnostic performance of FibroTouch. The area under the receiver operating characteristic curve for the LSM was 0.846 (95% confidence interval [CI] 0.808‐0.880) for fibrosis stage ≥ F1, 0.850 (95% CI 0.811‐0.883) for ≥ F2, 0.908 (95% CI 0.876‐0.934) for ≥ F3 and 0.874 (95% CI 0.836‐0.903) for F4. The diagnostic accuracy of LSM was superior to that of gamma‐glutamyl transpeptidase‐to‐platelet ratio (GPR), aminotransferase‐to‐platelet ratio index (APRI), or fibrosis index based on 4 factors (FIB‐4) index in staging fibrosis F2‐F4 (P = 0.007 to < 0.0001). Optimal LSM cut‐off values for diagnosing fibrosis stage ≥ F1, ≥ F2, ≥ F3, and F4 were 5.5 kPa, 7.85 kPa, 10.0 kPa, and 12.7 kPa, respectively.ConclusionFibroTouch has a high success rate and good reliability in staging liver fibrosis in patients with CHB.