Application of Fourier transform-based algorithms to resolve spectral overlapping for UV spectrophotometric co-assay of spiramycin and metronidazole in tablets

Doan Thanh Xuan,Vu Dang Hoang
DOI: https://doi.org/10.1016/j.saa.2022.121253
2022-09-05
Abstract:Fourier transform-based algorithms were investigated to resolve UV spectral overlapping of spiramycin and metronidazole in binary mixtures. UV spectra and ratio spectra were both subjected to fast Fourier transform-based first-order differentiation and discrete Fourier transform {i.e. using trigonometric functions such as sin xi - sin (xi + 450), cos xi + cos (xi + 450), sin 2xi - sin 2(xi + 450), cos 2xi + cos 2(xi + 450), sin xi - sin (xi + 600), cos xi + cos (xi + 600)} that followed by 3 passes of 17-point triangular smooth. Such signal transforms generated linear calibration graphs for either drug in the concentration range of 6.25 - 25 mg/L with R2 > 0.990. In comparison with the RP-HPLC reference method, the developed UV spectrophotometric methods had no significant difference in terms of accuracy and precision (p > 0.05) when quantifying spiramycin and metronidazole in their coated tablets. They are suggested as analytical quality control methods for their being environmentally friendly, technically simple, quick and economic.