Quantification of Antiviral Drug Tenofovir (TFV) by Surface-Enhanced Raman Spectroscopy (SERS) Using Cumulative Distribution Functions (CDFs)

Marguerite R. Butler,Jana Hrncirova,Meredith Clark,Sucharita Dutta,John B. Cooper
DOI: https://doi.org/10.1021/acsomega.3c07641
IF: 4.1
2024-01-16
ACS Omega
Abstract:Surface-enhanced Raman spectroscopy (SERS) is an ultrasensitive spectroscopic technique that generates signal-enhanced fingerprint vibrational spectra of small molecules. However, without rigorous control of SERS substrate active sites, geometry, surface area, or surface functionality, SERS is notoriously irreproducible, complicating the consistent quantitative analysis of small molecules. While evaporatively prepared samples yield significant SERS enhancement resulting in lower detection...
chemistry, multidisciplinary
What problem does this paper attempt to address?
The paper aims to address the reproducibility issues faced in quantitative analysis using Surface-Enhanced Raman Spectroscopy (SERS). Specifically, the paper addresses the following aspects: 1. **Non-reproducibility of SERS signals**: Due to the lack of strict control over the active sites, geometric structure, surface area, or surface functionalization of SERS substrates, SERS signals are unstable, making consistent quantitative analysis difficult. 2. **Challenges in quantitative analysis**: Although evaporative sample preparation can significantly enhance SERS signal intensity and lower the detection limit, these enhancement effects have a random spatial distribution on the SERS surface. 3. **Application of statistical methods**: To overcome these issues, the authors introduced a new analytical method, the cumulative distribution function (CDF), to process SERS data. This method improves the precision and accuracy of quantitative analysis. Through these methods, researchers were able to perform quantitative analysis of the antiviral drug Tenofovir (TFV) with a detection limit as low as 25 ng/mL, which is of significant importance in clinical applications. Additionally, the calibration curve obtained using CDF analysis demonstrated higher analytical sensitivity and lower reproducibility error compared to traditional SERS intensity calculation methods.