Patient‐ventilator interactions in new modes of patient‐triggered ventilation *

K. Abubakar,M. Keszler
DOI: https://doi.org/10.1002/ppul.1091
IF: 4.0899
2001-07-01
Pediatric Pulmonology
Abstract:Recently, synchronized modes of conventional mechanical ventilation became available for neonatal ventilatory support, but there has been little information regarding details of patient‐ventilator interactions during pressure support, volume support, or any other volume‐targeted modes of synchronized ventilation in newborn infants. Our objective was to obtain comparative data on patient‐ventilator interactions and stability of delivered tidal volume (VT) for the different modes of synchronized mechanical ventilation in stable ventilated newborn infants. We examined the effects of pressure support ventilation (PSV) and volume guarantee (VG) modes of a prototype Dräger Babylog® ventilator on peak and mean airway pressures (PIP and Paw), inspiratory time (tin), and VT in 23 ventilated newborn infants. Twelve infants were studied while on assist/control (AC) and 11 on synchronized intermittent mandatory ventilation (SIMV). Mean birth weight was 1,650 ± 1,180 g, gestational age 31 ± 6 weeks, and age at time of study was 19 ± 26 days. Data for 400–600 breaths from each infant were downloaded directly from the ventilator pressure and volume‐monitoring module, and analyzed using ANOVA for repeated measures. Mean values and breath‐to‐breath variability were compared for 20‐min periods of AC or SIMV followed by PSV, PSV+VG, and back to baseline AC or SIMV.