Standalone portable xenon-129 hyperpolariser for multicentre clinical magnetic resonance imaging of the lungs

Graham Norquay,Guilhem J Collier,Oliver I Rodgers,Andrew B Gill,Nicholas J Screaton,Jim Wild
DOI: https://doi.org/10.1259/bjr.20210872
2022-04-01
Abstract:Objectives: Design and build a portable xenon-129 (129Xe) hyperpolariser for clinically accessible 129Xe lung MRI. Methods: The polariser system consists of six main functional components: (i) a laser diode array and optics; (ii) a B0 coil assembly; (iii) an oven containing an optical cell; (iv) NMR and optical spectrometers; (v) a gas-handling manifold; and (vi) a cryostat within a permanent magnet. All components run without external utilities such as compressed air or three-phase electricity, and require just three mains sockets for operation. The system can be manually transported in a lightweight van and rapidly installed on a small estates footprint in a hospital setting. Results: The polariser routinely provides polarised 129Xe for routine clinical lung MRI. To test the concept of portability and rapid deployment, it was transported 200 km, installed at a hospital with no previous experience with the technology and 129Xe MR images of a diagnostic quality were acquired the day after system transport and installation. Conclusion: This portable 129Xe hyperpolariser system could form the basis of a cost-effective platform for wider clinical dissemination and multicentre evaluation of 129Xe lung MR imaging. Advances in knowledge: Our work successfully demonstrates the feasibility of multicentre clinical 129Xe MRI with a portable hyperpolariser system.