An Active Matrix Flat Panel Dosimeter (AMFPD) for in-phantom dosimetric measurements.

J. Moran,D. Roberts,T. Nurushev,L. Antonuk,Y. El‐Mohri,B. Fraass
DOI: https://doi.org/10.1118/1.1855012
2005-02-01
Abstract:An a-Si Active Matrix Flat Panel Imager (AMFPI) prototype developed in-house has been modified to function as an in-phantom dosimetry system providing high resolution two-dimensional (2-D) data. This Active Matrix Flat Panel Dosimeter (AMFPD) system can be used as a replacement device for standard in-phantom dosimeters, such as scanning ion chambers in water, or film in solid water. The initial characterization of the device demonstrates a wide dynamic range (up to 160 cGy), a stable calibration curve (less than 1.5% variation over 1 year), dose rate independence (less than 1%), and excellent agreement of output factors with ion chamber measurements for a range of field sizes (less than 2%). The device also compares well to film for 2-D planar dose distributions. It is expected that the AMFPD system will be useful for beam commissioning, algorithm verification test data, and routine IMRT quality assurance dosimetry.
What problem does this paper attempt to address?