Physical aspects of dynamic stereotactic radiosurgery.

Ervin B. Podgorsak,André Olivier,M. Pla,J. Hazel,A. Lotbinière,Bruce Pike
DOI: https://doi.org/10.1159/000100723
Applied Neurophysiology
Abstract:Dynamic stereotactic radiosurgery is a radiosurgical technique based on a medium-energy isocentric linear accelerator and a stereotactic frame. The technique uses concurrent and continuous rotations of both the gantry (300 degrees, from 30 to 330 degrees) and the couch (150 degrees, from 75 to -75 degrees). It gives a uniform dose (+/- 5%) within the target volume and dose fall-offs outside the target volume comparable to those obtained from presently known radiosurgical techniques.