Threshold-based Monitoring of Compound Muscle Action Potentials for Percutaneous Pedicle Screw Placement in the Lumbosacral Spine: Can We Rely on Stimulation of the Uninsulated Screw to Provide a Valid Safety Warning?

Yoichi Tani,Takanori Saito,Shinichiro Taniguchi,Masayuki Ishihara,Masaaki Paku,Takashi Adachi,Muneharu Ando,Nobuo Kohara,Jun Kimura
DOI: https://doi.org/10.1097/BRS.0000000000004263
2022-07-15
Abstract:Study design: A prospective analysis. Objective: To test if threshold-based monitoring of compound muscle action potentials (CMAPs) by stimulating the screw loaded to uninsulated extender sleeve provides a valid safety warning for percutaneous pedicle screw (PPS) placements in the lumbosacral spine. Summary of background data: Utility of the CMAP monitoring to PPS procedures remains controversial. Methods: A series of 202 patients underwent a total of 1664 lumbosacral PPS placements under CMAP monitoring without fluoroscopic guidance. The monitoring consisted of stimulating the PPS assembled to uninsulated extender sleeve and recording CMAPs from the vastus medialis, biceps femoris, tibialis anterior, and medial gastrocnemius. Automated steps of a threshold hunting algorithm using 0.2-ms duration pulses of increasing intensities delivered at 2/s allowed quick determination of a minimum stimulation current to evoke >100-μV amplitude CMAPs. Results: At L2 through S1 spines, postoperative CT scans identified 51 medial or inferior pedicle wall breaches of 1536 screws (3.3%) without neurologic complications. The receiver operating characteristic curve analysis determined the critical cutoff threshold value of 27 mA (74% sensitivity and 95% specificity) for predicting 35 breaches of 627 screws (5.6%) at L2 and L3, and of 17 mA (100% sensitivity and 98% specificity) for 16 of 909 (1.8%) at L4 through S1. While advancing the screw, three breaches (5.9%) showed a particularly low threshold of ≤6-mA, allowing the surgeon to immediately redirect the screw and retest the new trajectory as safe. Conclusion: Screw stimulation with threshold hunting algorithm has a distinct advantage over the time-consuming insulated pilot hole stimulation, allowing an uninterrupted flow of the surgery. The present findings have documented practical usefulness and reliability of CMAP monitoring using direct stimulation of the PPS assembled to uninsulated extender sleeve.
What problem does this paper attempt to address?