High-Resolution Qualitative and Quantitative Magnetic Resonance Evaluation of the Glenoid Labrum
K. Iwasaki,Monica Tafur,E. Chang,S. Statum,R. Biswas,B. Tran,W. Bae,J. Du,G. Bydder,C. Chung
DOI: https://doi.org/10.1097/RCT.0000000000000307
2015-11-01
Journal of Computer Assisted Tomography
Abstract:Objective This study aimed to implement qualitative and quantitative magnetic resonance sequences for the evaluation of labral pathology. Methods Six glenoid labra were dissected, and the anterior and posterior portions were divided into normal, mildly degenerated, or severely degenerated groups using gross and magnetic resonance findings. Qualitative evaluation was performed using T1-weighted, proton density-weighted, spoiled gradient echo and ultrashort echo time (UTE) sequences. Quantitative evaluation included T2 and T1rho measurements as well as T1, T2*, and T1rho measurements acquired with UTE techniques. Results Spoiled gradient echo and UTE sequences best demonstrated labral fiber structure. Degenerated labra had a tendency toward decreased T1 values, increased T2/T2* values, and increased T1rho values. T2* values obtained with the UTE sequence allowed for delineation among normal, mildly degenerated, and severely degenerated groups (P < 0.001). Conclusions Quantitative T2* measurements acquired with the UTE technique are useful for distinguishing among normal, mildly degenerated, and severely degenerated labra.