Contrast-Enhanced Ultrasonography as a Novel Method for the Dynamic Visualization of Blood Flow and Fiber Blockage in Dialyzers: A Feasibility Study

Franz Josef Putz,Ernst Michael Jung,Christina Putz,Miriam C Banas,Tobias Bergler,Jörg Vienken,Bernhard Banas
DOI: https://doi.org/10.1016/j.ultrasmedbio.2020.05.005
Abstract:The capillary dialyzer represents the central element of the extracorporeal blood circuit of a therapy system for hemodialysis. The aim of this study was to assess the blood-flow characteristics of dialyzers with the help of modern ultrasound techniques. Five brand-new dialyzers (FX80 classix, Fresenius Medical Care, Bad Homburg, Germany) and five dialyzers after a dialysis session were analyzed by different ultrasound techniques to detect functional and structural changes. B-mode and Doppler techniques were not suitable to describe differences in brand-new and clinically applied dialyzers. Contrast-enhanced ultrasonography, however, was able to visualize blood-flow profiles in the capillaries. Although dialyzers displayed no signs of clinical dysfunction, contrast-enhanced ultrasonography was able to detect blocked capillaries of varying degrees after a dialysis session in all five examined dialyzers. Consequently, the blood-flow velocity was higher in the remaining unblocked capillaries in comparison to the velocity in the brand-new dialyzers. This information may be helpful for improving the geometric design of dialyzers, including their capillary membranes, and optimizing anti-coagulation strategies in hemodialysis patients.
What problem does this paper attempt to address?