Assessment of Pull-out Forces in TEVAR and ANACONDA FEVAR Combination and Early Clinical Results: Creation of a Proximal Landing Zone for FEVAR in Patients with Extent I and Extent IV TAAAs

Jürgen Falkensammer,Fadi Taher,Markus Plimon,Miriam Kliewer,Corinna Walter,Elisabeth Pelanek,Afshin Assadian
DOI: https://doi.org/10.1016/j.avsg.2020.01.075
Abstract:Background: Although recent data on the treatment of thoracoabdominal aortic aneurysms (TAAAs) are promising, in some cases, the paravisceral segment of the aorta may not be suitable for a branched endograft due to space restrictions. A combination of a fenestrated aneurysm repair (FEVAR) with a thoracic aneurysm repair (TEVAR) may represent a feasible treatment option. The current investigation was performed to assess the stability of a fenestrated Anaconda device implanted into a set of thoracic endografts from different manufacturers. We then assessed our clinical results with the FEVAR/TEVAR combination. Methods: Pull-out forces were measured in vitro after docking a fenestrated Anaconda graft within the distal end of different TEVAR devices. Anaconda devices were implanted in 28- or 30-mm thoracic tube grafts (oversizing of at least 2 mm: 13.3-21.4; minimum overlap of 15 mm). Continuously increasing longitudinal pull forces of up to 100 N were applied on an Instron Tensile Tester. Initial break point and damage to the endografts were documented. Clinical results of patients treated with such an FEVAR/TEVAR combination at our institution are presented as a second part of this study. Results: Median pull-out forces ranged from 2.38 N to 55.0 N. The highest stability was achieved with 34-mm Anaconda devices in 28-mm thoracic tube grafts. Grafts with either thinner Dacron material or those featuring a polytetrafluorethylene membrane seemed especially vulnerable to punctures and tears caused by the downward-looking hooks of the Anaconda device. Between April 1, 2013 and December 31, 2018, in 28 of 172 patients treated with a fenestrated Anaconda device, prior implantation of a thoracic tube graft was necessary to create a sufficient proximal landing zone. In 25 cases (89.3%), the aneurysm was successfully treated. Although the 30-day reintervention rate in this subgroup was relatively high at 28.6%, none of these was due to a failure of the FEVAR/TEVAR combination. Upon an average follow-up of 15 months, no failure of the graft connection and no type III endoleak due to membrane damage were observed. Conclusions: The combination of a thoracic tube graft and a fenestrated Anaconda device is a viable option for the treatment of patients with Extent I or IV TAAAs with no adequate landing zone above the celiac trunk. Although pull-out testing has shown good stability with most assessed grafts, the thoracic devices with thicker Dacron membranes seemed to be especially suitable.
What problem does this paper attempt to address?