Measurement of ceruloplasmin from its oxidase activity in serum by use of o-dianisidine dihydrochloride.

K. Schosinsky,H Peter Lehmann,M. Beeler
DOI: https://doi.org/10.1093/CLINCHEM/20.12.1556
IF: 12.114
1974-12-01
Clinical Chemistry
Abstract:We have developed a new method for determination of serum ceruloplasmin from its oxidase activity, in which o -dianisidine dihydrochloride is used as substrate. o -Dianisidine dihydrochloride is more stable than is the more widely used p -phenylenediamine substrate, and forms a stable product that is measured at 540 nm. Correlation was good between results of this method and those obtained with both a p -phenylenediamine substrate method ( r = 0.98) and a radial immunodiffusion procedure ( r = 0.97). There is little or no interference from reducing or colored components of serum. The coefficient of variation (day-to-day) for the method was 4.2%. A normal range of 62-140 U/ liter has been determined by the reported method.
What problem does this paper attempt to address?