Tracking accuracy evaluation of electromagnetic sensor-based colonoscope tracking method

Masahiro Oda,Hiroaki Kondo,Takayuki Kitasaka,Kazuhiro Furukawa,Ryoji Miyahara,Yoshiki Hirooka,Hidemi Goto,Nassir Navab,Kensaku Mori
DOI: https://doi.org/10.48550/arXiv.2201.07167
IF: 4.506
2022-01-14
Medical Physics
Abstract:This paper reports a detailed evaluation results of a colonoscope tracking method. A colonoscope tracking method utilizing electromagnetic sensors and a CT volume has been proposed. Tracking accuracy of this method was evaluated by using a colon phantom. In the previously proposed paper, tracking errors were measured only at six points on the colon phantom for the accuracy evaluation. The point number is not enough to evaluate relationships between the tracking errors and positions in the colon. In this paper, we evaluated the colonoscope tracking method based on more detailed measurement results of the tracking errors. We measured tracking errors at 52 points on the colon phantom and visualized magnitudes of the tracking errors. From our experiments, tracking errors in the ascending and descending colons were enough small to perform colonoscope navigations. However, tracking errors in the transverse and descending colons were large due to colon deformations.
What problem does this paper attempt to address?