Abstract:BackgroundThe Point of Care (POC) department was notified by an end user in the cardiac catheterization laboratory (Cath Lab) of i-STAT activated clotting time (ACT) results not transmitting to a patient's chart. Subsequent audits determined that this failed transmission was due to a delay in properly docking of the i-STAT into its downloader. Upon a broader and more thorough investigation, it became clear that this was a systemic issue among und-users of all i-STAT ACT testing for heparin dosage monitoring, as many cases went without transmission to patient electronic medical records (EMR). Not only did this present patient safety concerns related to the lack of documentation of test performance, it led to an approximate 30,000.ConclusionsThis project shows that open lines of communication to patient care areas combined with timely audits by POC Coordinators can have tremendous benefits on patient care at little to no additional institutional cost. Periodic POC quality assurance should include review of successful connectivity for interfaced test systems to ensure all testing performed is documented to the EMR.