Validation of a Low‐Cost Manometer to Assess of Tongue, Lip, Cheek, and Respiratory Strength: A Laboratory‐Based Study

James A. Curtis,Crystal Diaz,Theresa Lee,Anaïs Rameau
DOI: https://doi.org/10.1002/lary.31727
IF: 2.97
2024-08-26
The Laryngoscope
Abstract:Findings from this study suggest that the low‐cost generic pressure testing manometer used in this study has the potential to be a valid substitute for the IOPI and MicroRPM for measurement of lingual, facial, and respiratory muscle strength during clinical assessments of speech, cough, and swallowing. Objective(s) The objective of this study was to characterize the level of agreement between three manometers: (1) Iowa Oral Performance Instrument (IOPI)—the reference standard for tongue, lip, and cheek strength assessments; (2) MicroRPM Respiratory Pressure Meter (MicroRPM)—the reference standard for respiratory strength assessments; and (3) Digital Pressure Manometer (DPM)—an alternative, low‐cost pressure testing manometer. Methods Manual pressures were simultaneously applied to the IOPI and DPM, and to the MicroRPM and DPM, within a controlled laboratory setting. Agreement in pressure readings were analyzed using descriptive statistics, Lin's concordance correlation, and Bland–Altman Plots. Agreement was interpreted as "poor" if ρc
medicine, research & experimental,otorhinolaryngology