DOI: https://doi.org/10.1016/j.aca.2003.10.041
IF: 6.911
2004-02-01
Analytica Chimica Acta
Abstract:A quantitative liquid chromatographic–electrospray ionization mass spectrometric method for the determination of dexamethasone in sheep plasma has been developed and validated. The samples were extracted using solid-phase extraction cartridges with mixed reversed-phase materials (oasis-HLB). The chromatographic separation was performed on a reversed-phase XTerrra MS C18 column (3.0mm×150mm; 5μm) using a mobile phase consisting of 65% methanol in water containing 0.1% (v/v) formic acid, pumped at a flow rate of 0.30mlmin−1. The analyte was detected after positive electrospray ionization using selected ion monitoring (SIM) mode. The probe heater temperature was set at 260°C, the capillary voltage was set at 3.5kV and the source block voltage (AQAmax) was set at 30V. The method was fully validated. Calibration graphs were linear (r better than 0.998, n=11), in concentration ranges 6–1000ngml−1 for dexamethasone. The intra- and inter-day RSD values were less than 24.1% (n=6). The limits of detection and quantitation for dexamethasone were found to be 1 and 6ngml−1, respectively. The efficiency of the solid phase extraction procedure was found to be 92.4% for dexamethasone. The method was further applied to a pilot kinetic study in order to assess the main pharmacokinetic parameters of dexamethasone in sheep.