Pre-processing of Photoplethysmographic Waveform for Amplitude Regularization

Jaepil Kim,Jeong-Whan Lee,Hangsik Shin
DOI: https://doi.org/10.1007/s42835-019-00185-y
2019-05-29
Journal of Electrical Engineering and Technology
Abstract:This study is aimed to develop a post-processing method to regularize amplitude of PPG waveforms and evaluate the performance of the algorithm. The proposed procedure for regularizing amplitude consists of extracting the signal envelopes based on time-domain filtering, generating a compensation curve, and reconstructing the waveform. The proposed technique was evaluated with respect to dispersion and waveform fluctuation by means of coefficients of variation (CV) and a scaling exponent of detrended fluctuation analysis (DFA). Other criteria included the detection performances of major features such as upper and lower fiducial points and maximum dV/dt. In developing this method, we used a clinical dataset that comprised 74,225 beats. By applying reconstruction algorithm, the %CV and the scaling exponent of DFA were significantly decreased from 115.4 to 71.4 and from 1.029 to 0.836, respectively in peak amplitude. In valley amplitude, they were also decreased from 38.3 to 13.2 and from 0.993 to 0.713, respectively. Moreover, sensitivity of maximum dV/dt point, upper and lower peak detection are increased by 8.50, 2.72 and 5.21% after reconstruction, respectively. Our results clearly showed that the proposed reconstruction technique could reduce amplitude fluctuation and improve detection of fiducial points.
engineering, electrical & electronic
What problem does this paper attempt to address?