The CRILIN calorimeter: gamma radiation resistance of crystals and SiPMs

A. Cemmi,B. D'Orsi,E. Di Meco,I. Di Sarcina,E. Diociaiuti,M. Moulson,D. Paesani,I. Sarra,J. Scifo,A. Verna
DOI: https://doi.org/10.1088/1748-0221/19/10/p10016
2024-10-17
Journal of Instrumentation
Abstract:The Crilin calorimeter is a semi-homogeneous calorimetric system based on Lead Fluoride (PbF 2 ) crystals with UV-extended Silicon Photomultipliers (SiPMs) proposed for the Muon Collider. This study investigates the radiation resistance of crystals and SiPMs, subjected to 10 kGy gamma irradiation, equivalent to a 10-year service life in the Muon Collider. Our findings indicate that while PbF 2 crystals exhibit a decrease in transmittance post-irradiation with partial recovery over time, the alternative PbWO 4 -Ultra Fast (PWO-UF) demonstrates exceptional radiation hardness, maintaining stable transmittance. SiPMs showed an increase in dark current and breakdown voltage post-irradiation, with less degradation observed in the SiPM biased during the exposure to radiation compared to the unbiased component. These results underscore the viability of PbF 2 for radiation-tolerant calorimeters, though improvements in production homogeneity are needed. The superior performance of PWO-UF crystals suggests they are a promising alternative for high-radiation applications, but their higher cost must be carefully considered.
instruments & instrumentation
What problem does this paper attempt to address?