Development and validation of a sensitive LC-MS/MS method for simultaneous determination of eight tyrosine kinase inhibitors and its application in mice pharmacokinetic studies

Yu He,Lei Zhou,Song Gao,Taijun Yin,Yifan Tu,Robert Rayford,Xiaoqiang Wang,Ming Hu
DOI: https://doi.org/10.1016/j.jpba.2017.09.013
2018-01-30
Abstract:The purpose of this study was to develop and validate a robust and sensitive LC-MS/MS method for simultaneous determinations of various tyrosine kinase inhibitors (TKIs) in biological samples and to apply the method to their pharmacokinetic studies. Processed samples were injected into the UHPLC system coupled to an ESI-triple quadrupole mass spectrometer. The compounds were separated on an AcQuity UHPLC BEH C18 column (50mm×2.1mm ID, 1.7μm) using a gradient elution of acetonitrile/0.1% formic acid in water. The mass analysis was performed in an API 3200 Qtrap mass spectrometer via selective reaction monitoring operated under a positive scanning mode. The method was validated over a linear range of 3.13-800nM for erlotinib; 6.25-1600nM for sunitinib, pazopanib, and axitinib; and 12.5-3200nM for sorafenib, dasatinib, lapatinib, and nilotinib, respectively. The intra-day and inter-day precision were <16.7% for quality control samples of the analytes at the low concentration level and <13.7% for all other concentrations. The accuracy (bias) for all analytes at three different concentration levels ranged from -12.2% to 15.0%. The recovery, matrix effect, and stability were all in the range of acceptance. Only 10μl of blood were needed, demonstrating the method's high sensitivity. The presented method was shown to be suitable for the analysis of serial blood samples collected from each mouse in a pharmacokinetic study, after the oral administration of 11 TKIs (each at 1mg/kg) as a mixture.
What problem does this paper attempt to address?