A clinical trial on a plate? The potential of 384-well format solid phase extraction for high-throughput bioanalysis using liquid chromatography/tandem mass spectrometry.

R. Biddlecombe,C. Benevides,S. Pleasance
DOI: https://doi.org/10.1002/1097-0231(20010115)15:1<33::AID-RCM188>3.0.CO;2-4
IF: 2.586
2001-01-15
Rapid Communications in Mass Spectrometry
Abstract:The application of 384-well format solid phase extraction (SPE) for bioanalysis using liquid chromatography/tandem mass spectrometry (LC/MS/MS) is reported and a 384-well SPE method for the 5-HT agonist sumatriptan in human plasma described. Plasma samples were extracted on a prototype low-density polyethylene 384-well SPE block using a packed bed of 5 mg Oasistrade mark HLB. Liquid handling was automated by a combination of a robotic sampler processor and a 96/384 multi-channel dispensing station. Samples and SPE reagents were drawn through the SPE block by centrifugation. The extracts were analysed by LC/MS/MS with thermally and pneumatically assisted electrospray ionisation and selected reaction monitoring. The method is used to illustrate and discuss the feasibility and viability of sample preparation techniques in high-density microtitre plate format for routine bioanalysis.
What problem does this paper attempt to address?