Radiation exposure from EMI scanner-multiple scans.

R. Horsley,V. G. Peters
DOI: https://doi.org/10.1259/0007-1285-49-585-810
1976-09-01
British Journal of Radiology
Abstract:The measurement of radiation exposure to the skull for a single complete scan has been reported by Perry and Bridges (1973). They conclude that the exposure to the cranium for a complete series of transverse axial scans is between 1.0 and 2.5 R. This result follows from the assumption that there is a negligible contribution to the dose due to scattered radiation from adjacent scans. Indeed, the narrow, highly collimated beam of X rays should exhibit narrow beam characteristics with a backscatter factor approaching unity. However, as the beam scans back and forth moving in an arc around the head, and as additional scans are performed at different levels, the area irradiated by the X-ray beam becomes considerably larger. The size of the collimated beam is about 2.5 cm × 0.3 cm, but the actual size of the field over a single traverse of the X-ray beam is 2.5 cm × 25.6 cm. The routine skull examination includes three or four complete scans performed at 2.5 cm intervals. Thus the total field size for a full se...