This is the first in‐human trial and prospective case series of a novel single‐port robotic system for gynaecological surgery: An IDEAL stage 2a study

Xueli Hu,Mengjia Ruan,Luoyi Zhu,Menghan Huang,Lifang Qi,Min Huang,Shuangwei Zou,Lili Lin,Wenwen Zheng,Lizhi Wang,Ping Duan
DOI: https://doi.org/10.1002/rcs.2657
2024-09-22
International Journal of Medical Robotics and Computer Assisted Surgery
Abstract:Abstract Background The Shurui® system (SR‐ENS‐600) is a novel fully integrated single‐port robotic system with bioinspired serpentine surgical manipulators and a camera. Methods This was a single‐centre prospective case‐series study according to the IDEAL stage 2a guidelines to evaluate the feasibility, safety and potential efficacy of the Shurui® system for gynaecological surgeries and to improve the operating process. Results Ten patients with a gradient of surgical difficulty who had indications for laparoscopic surgery and who volunteered to participate in a clinical trial were enrolled in the study. All 10 subjects successfully completed the procedure without converting to other procedures. No serious complications were reported at the 3‐month follow‐up. Subjects recover faster after surgery and are highly satisfied with the incision. Conclusions Gynaecological single‐site laparoscopic surgery with the Shurui® system was technically feasible for well‐selected patients with minimal alterations in technique. Further prospective multicenter large‐sample studies are necessary. Registration number ChiCTR2300075431. URL: https://www.chictr.org.cn/showproj.html?proj=189995 .
surgery
What problem does this paper attempt to address?