Competitive immunoassay using enzyme-regulated Fe3O4@COF/Fe3+ fluorescence probe for natural chloramphenicol detection

Jutao Hu,Xiaoyu Su,Lei Yuan,Kaiyi Zheng,Xiaobo Zou,Zongbao Sun,Xuechao Xu,Wen Zhang
DOI: https://doi.org/10.1016/j.aca.2023.341680
IF: 6.911
2023-07-01
Analytica Chimica Acta
Abstract:Accurate and sensitive detection of chloramphenicol (CAP) in natural samples is essential for ensuring human health. Herein, an enzyme-regulated fluorescence sensor using Fe<sub>3</sub>O<sub>4</sub>@COF/Fe<sup>3+</sup> probe, is developed for CAP determination. Fe<sub>3</sub>O<sub>4</sub>@COF, synthesized via hydrothermal method, exhibits dual functions as a magnetic carrier and signal probe. Bovine serum albumin conjugated-chloramphenicol, adsorbed on the surface of Fe<sub>3</sub>O<sub>4</sub>@COF, competes with CAP for antibody binding. The antibody interacts with alkaline phosphatase via the biotin-streptavidin system. Meanwhile, ascorbic acid, produced from the enzyme-catalyzed reaction dominated by alkaline phosphatase, effectively restores the fluorescence of Fe<sub>3</sub>O<sub>4</sub>@COF that is quenched by Fe<sup>3+</sup>. After experimental verification and gradual optimization, a logarithmic linear relationship between CAP concentration and fluorescence intensity is established in the range of 2 × 10<sup>-4</sup>∼10 μg mL<sup>-1</sup>, with a good limit of detection (9.2 × 10<sup>-5</sup> μg mL<sup>-1</sup>). Proposed method exhibits excellent stability (15 days) and reusability (8 cycles), providing a sensitive and reliable method for accurate CAP detection. The readouts show good agreement with HPLC and recoveries during laboratory and natural CAP analysis.
chemistry, analytical
What problem does this paper attempt to address?