Evaluation of a New Immunochromatographic Assay Kit for the Rapid Detection of Norovirus in Fecal Specimens

Kwi Sung Park,Kyoung Ah Baek,Dong Uk Kim,Kyung Sook Kwon,Sun Hye Bing,Joon Soo Park,Hae Seon Nam,Sang Han Lee,Young Jin Choi
DOI: https://doi.org/10.3343/alm.2012.32.1.79
IF: 4.941
2012-01-01
Annals of Laboratory Medicine
Abstract:Rapid and accurate detection of norovirus is essential for the prevention and control of norovirus outbreaks. This study compared the effectiveness of a new immunochromatographic assay kit (SD BIOLINE Norovirus; Standard Diagnostics, Korea) and real-time reverse transcription-PCR (RT-PCR) for detecting norovirus in fecal specimens. Compared with real-time RT-PCR, the new assay had sensitivity, specificity, positive predictive value, and negative predictive value of 76.5% (52/68), 99.7% (342/343), 98.1% (52/53), and 95.5% (342/358), respectively. The sensitivity of the assay was 81.8% (18/22) for GII.3 and 75.7% (28/37) for GII.4. None of the 38 enteric virus-positive specimens (3 for astrovirus, 5 for enteric adenovirus, and 30 for rotavirus) tested positive in the cross-reactivity test performed by using this assay. The new immunochromatographic assay may be a useful screening tool for the rapid detection of norovirus in sporadic and outbreak cases; however, negative results may require confirmatory assays of greater sensitivity.
medical laboratory technology
What problem does this paper attempt to address?