Evaluation of Coincidence Time Resolution in a liquid xenon detector with silicon photomultipliers

N. Salor-Iguiñiz,J.M. Benlloch-Rodríguez,R. Esteve,C. Romo-Luque,R.J. Aliaga,V. Álvarez,F. Ballester,R. Gadea,F. Monrabal,M. Querol,M. Rappaport,J. Rodríguez,J. Rodríguez-Ponce,S. Teruel-Pardo,J.F. Toledo,R. Torres-Curado,P. Ferrario,V. Herrero,J.J. Gómez-Cadenas
2024-12-09
Abstract:This work explores the combination of liquid xenon as a scintillating medium and silicon photomultipliers as a readout in Positron Emission Tomography (PET) for enhanced Time-Of-Flight resolution. We present the results of our first prototype optimized to maximize light collection using high-PDE, VUV-sensitive sensors and to minimize time fluctuations. We report a coincidence time resolution of 281 $\pm$ 2 ps FWHM, obtained using a $^{22}$Na calibration source. This result is competitive with the current state-of-the-art PET scanners and represents a significant step forward in the development of liquid xenon as a viable alternative to conventional scintillators in PET technology.
Instrumentation and Detectors,Medical Physics
What problem does this paper attempt to address?